This machine is a testing instrument used to measure the concentration of pencil lead. Specifically, it measures the reflection optical density (represented by the symbol d: d=Lg1/R450, where R450 is the diffuse reflectance brightness factor of the sample for blue light with a wavelength of 450nm, that is, under certain illumination conditions and in a given direction, the ratio of the reflected light brightness of the sample to the ideal diffuse reflectance brightness under the same conditions) of the lead traces scribed on a specified sample paper.


Product Detail

Product Tags

Product Parameters
1 Elimination of specular reflection The error is less than 0.1% of the brightness factor, having no impact on concentration measurement.
2 Measurement range 0≤d≤2
3 Sample size 25mm square
4 Measurement hole circular with a diameter of 22mm
5 Repeatability Brightness factor error less than 0.1%
6 Indication error Luminance factor error less than 0.3%
7 Power supply 220V±10%, 50Hz
8 Working environment Temperature 10~30℃, relative humidity ≤85%
9 Overall dimensions (length * width * height) 365 * 245 * 430mm
10 Instrument weight approximately 11kg

 

Standard
Comply with the relevant requirements of standard 5.5 in QB/T2774-2006.

 