LT-CJ80 Cold And Hot Impact Test Chamber
| 1 | Size of the inner box (W * H * D) | 500 * 400 * 400mm (80L) |
| 2 | Size of the outer box (W * H * D) | 1480 * 1900 * 1550mm |
| 3 | Temperature range | (-40℃, -50℃, -60℃) ~ + 150℃ |
| 4 | Temperature range in the high temperature area | + 60℃ ~ + 150℃ |
| 5 | Temperature range of the test area | + 150℃ ~ -60℃ |
| 6 | Temperature range in the low-temperature area | -10℃ ~ -70℃ |
| 7 | Test method | pneumatic valve switch 2 temperature area or 3 temperature area |
| 8 | Temperature shock range | -60~ + 150℃ |
| 9 | Temperature fluctuation degree | ± 0.5℃ |
| 10 | Temperature deviation | ± 2.0℃ |
| 11 | Temperature recovery time | 3~5 minutes |
| 1 | Electrical control system | a. The controller adopts color touch screen Korean 8226S Chinese and English display + PLC (control software) + temperature control module; b. Operation mode: program mode. |
| 2 | Heating and air supply air circulation system | a. Working mode: mechanical compression binary folding refrigeration mode; b. Refrigeration compressor: original imported fully enclosed piston compressor; c. Evaporator: fin-tube heat exchanger; d. Condenser: bushing-type heat exchanger; e. Saving device: thermal expansion, capillary; f. Evaporative condenser: stainless steel brazed plate type heat exchanger; g. Chiller control mode: PLC controls the operating conditions, compressor return gas cooling circuit and energy adjustment loop according to the test conditions; h. Refrigerant: R404A / R23; ozone depletion index is 0; i. Others: Compressor cooling fan according to the size of compressor. |
| 3 | Heating and air supply circulation system: | a. Using multi-wing centrifugal circulation fan, reinforced shaft + high and low temperature resistance of rotating blade aluminum alloy, to reach forced convection; b. High temperature / ring temperature exposure fan: 3,750W centrifugal type; low temperature exposure fan; 3,750W centrifugal type; c. Nickel-chromium alloy electric heating wire heater; d. Heater control mode: no-contact periodic pulse broadening, SSR, solid state relay). |